Symptom Potential Cause Recommended Troubleshooting/Test
Remote not powering on/ Screen is blank Battery is not charged Connect remote to charger &  allow to fully charge
Device malfunctioning Contact Compliant Technologies
E-Band not powering on/ No lights illuminated Battery is not charged Connect remote to charger &  allow to fully charge
Device not powered on with magnetic key Utilize magnetic key to power on device to STANDBY mode
Magentic key damaged Attempt to power on with a known good magnetic key
Device malfunctioning Contact Compliant Technologies
Remote not charging Poor connection to USB cable Ensure firm connection between remote & cable & between cable & wall charger cube
USB cable damaged/faulty Attempt charging with a known good cable
Wall charger cub damaged/faulty Attempt charging with a known good charger cube
Battery life exceeded Contact Compliant Technologies
E-Band not charging Poor connection to USB cable Ensure firm connection between remote & cable & between cable & wall charger cube
USB cable damaged/faulty Attempt charging with a known good cable
Wall charger cub damaged/faulty Attempt charging with a known good charger cube
Battery life exceeded Contact Compliant Technologies
Stimulation ineffective/non-functional Poor connection to subject Ensure good electrode contact to subject's skin (reposition, eliminate hair, etc)

Device malfunction  Perform Tap Test
 Contact Compliant Technologies

Remote malfunction Perform Tap Test using known- good remote to stimulate the device



To Initiate Audible Warning: On the remote press 4 and OK. Audible warning will continue until disabled.

To Disable Audible Warning: On the remote press 3 and OK